To clean your Device, follow these step-by-step instructions:
- Step 1 - Power Down: Ensure the Device is powered off and cooled down.
- Step 2 - Remove Fit Kit accessories: Remove the Nose Pad, Forehead Pad, and lens insert, if applicable.
- Step 3 - Using a wipe from these compatible products: Super Sani Cloth or Cavi Wipes to perform step 4.
Note: Prior to using each new wipe, squeeze the excess fluid out of the wipe so that it is damp but not dripping.
Caution: Do not spill or oversaturate any system seams, ports, or perimeter of the
lenses.
Caution: Do not use abrasive products such as paper towels. - Step 4 - Cleaning: Wipe all surfaces of the Device (including disassembled pieces and cables) to remove any dirt, debris, or residue.
Note: Use additional wipes as needed, remember to squeeze the excess fluid out of the wipe so that it is damp but not dripping. - Step 5 - Disinfection: Using a new wipe, clean all surfaces of the Device (including disassembled pieces and cables) and ensure the proper disinfection contact time per label of the wipe. Use additional wipes as needed to maintain contact time. (the contact time for both Super Sani cloths and CaviWipes is 2 minutes)
- Step 6 - Rinse the Device by wiping it with an isopropyl alcohol wipe or clean sterile lint free cloth dampened (not dripping) with isopropyl alcohol. Ensure to gently wipe over the cameras and display lenses of the headset and cameras of the control.
Note: Gently wipe over the cameras and display lenses of the headset and cameras of the control. - Step 7 - Rinse the Device by wiping it with a clean, sterile, lint-free cloth dampened (not dripping) with utility (tap) water.
- Step 8 - Dry the Device by wiping it with a sterile lint free cloth